Spec sheet of Varian GC/MS @ AECL
Injector: 1177 type
Column: CP-Select 624CB
GC: 3800
Detector: model 300 single-quadrupole MS.
The EI source consists of an ion volume, a filament assembly, electron collimating magnets, and ion focusing lenses, all supported by a heated ion block. The ion volume is an open cylinder with two side holes. Vaporized sample and carrier gas from the GC enters the ion volume through the transfer line in one of the holes. An electron beam, generated at the heated filament, enters the ion volume through the other hole. The accelerated electrons collide with the sample molecules inside the ion volume and generate molecular ions. The resulting ions fragment into differently charged or neutral fragments. The ions enter the mass analyzer [17].
Air water check
Expected
28 m/z Size < 20.0 mV
28/32 m/z Ratio < 3.5:1
28/18 m/z Ratio < 1.2:1
MS Drift
Expected
EI Vacuum < 30.0 mTorr
Manifold Vacuum < 1×10-6 Torr
Lens 1 Voltage > -15V